Interface 1600 Gold Standard Calibration Load Cell

The 1600 Gold Standard series is a calibration grade Low Profile load cell that offers improved accuracy of performance and temperature characteristics. Additional gauges help improve the performance and the locked-in calibration adaptor removes uncertainty of thread engagement which with a stiff flexure could otherwise cause poor non-repeatability at lower applied forces. Every unit is supplied with a NIST traceable ASTM E74 or the option of a ISO376 calibration. Both calibrations include traceable 3-run calibration certificate that includes graphical plots and calculations of figures for polynomial fit equation. The calibration starts from just 4% of capacity.

  • 0.005% non-repeatability
  • 0.01% Creep
  • High output - to 4mV/V
  • 3 run NIST traceable ASTM E74 Calibration
  • 4% lower load limit
  • Optional ISO376 calibration
  • Universal or Compression only options
  • Warranted performance to 0.02%
  • Warranted non-repeatability to +/- 0.005% or better
  • 9 capacities from 2.2kN to 900kN (500lbf to 200Klbf) High output to 4.0 mV/V for highest resolution.
  • Fitted with locked-in calibration adaptor
  • Moment compensation for enhanced insensitivity to extraneous load.
  • Excellent temperature characteristics
  • Low deflection
  • Individual 3-run multi-point calibration certificate traceable to National Standards
PARAMETERSMODEL
16101610161016201632, 1640
CAPACITY
Measuring RangeU.S. (lbf)5K, 10K1K, 2K50025K, 50K100K, 200K
Metric (kN)22, 454.5, 92.2110, 225450, 900
ACCURACY - (MAX ERROR)
Static Error Band - %FS±0.025±0.025±0.02±0.05±0.05
Nonlinearity - %FS±0.035±0.035±0.03±0.05±0.05
Hysteresis - %FS±0.045±0.035±0.02±0.05±0.05
Non-repeatability - %RO±0.005±0.005±0.005±0.005±0.005
Creep, in 20 min - %±0.01±0.01±0.01±0.01±0.01
Side Load Sensitivity - %±0.1±0.1±0.1±0.1±0.1
Eccentric Load Sensitivity - % / in±0.1±0.1±0.1±0.1±0.1
TEMPERATURE
Compensated Range°F+14 to +113+14 to +113+14 to +113+14 to +113+14 to +113
°C-10 to +45-10 to +45-10 to +45-10 to +45-10 to +45
Operating Range°F-67 to +194-67 to +194-67 to +194-67 to +194-67 to +194
°C-55 to +90-55 to +90-55 to +90-55 to +90-55 to +90
Effect on Zero - %RO MAX°F±0.0008±0.0008±0.0008±0.0008±0.0008
°C±0.0014±0.0014±0.0014±0.0014±0.0014
Effect on Output - %RO MAX°F±0.0008±0.0008±0.0008±0.0008±0.0008
°C±0.0014±0.0014±0.0014±0.0014±0.0014
ELECTRICAL
Rated Output - mV/V (Nominal)44244
Bridge Resistance - Ohm (Nominal)350350350350350
MECHANICAL
Safe Overload - %CAP±150±150±150±150±150
Deflection @ ROin0.0040.0040.0020.0060.01
mm0.1020.1020.0510.1520.254
Optional Base - P/N (Metric)installed_baseinstalled_baseinstalled_baseinstalled_baseinstalled_base
Weightlb23.583.858171
kg10.663.631.7226.3177.56
MaterialAluminumAluminumAluminumAlloy SteelAlloy Steel
CalibrationTension & Compression

  • Multiple independent bridge outputs
  • Special fittings
  • Integral cable or choice of connector(s) Mechanical adaptations
  • Standardised output
  • A2LA accredited calibration
  • ID Chip or TEDS (Transducer Electronic Data Sheet)
  • ISO/ASTM or MILSPEC calibration
  • Special thread sizes
  • Display electronics or data acquisition
  • Load cell with fitted base and locked-in calibration adaptor
  • Connector output versions supplied with mating part(s)
  • NIST traceable, ASTM E74 or ISO376 calibration certificate with plots and calculations

Why choose Interface Sensor Products?

Interface is an internationally recognised brand and decades of experience have gone into making Interface manufactured load cells the benchmark for thousands of users around the world.

Interface product reliability is second to none and all our product lines have to match this standard before we would consider recommending them to our customers.

Our customer list is a "Who's Who" of household names ranging from the Aerospace Giants through to the most prestigious Universities. This speaks volumes about our reputation for quality solutions on projects of all sizes around the world.

Back To Top