Interface 1600 Gold Standard Calibration Grade Low Profile Load Cell

The Interface 1600 Gold Standard Calibration Grade Low Profile Load Cell is an industry-leading reference force sensor custom-engineered specifically to calibrate other secondary load cells and audit high-precision materials testing frames. Engineered to deliver a significant step up from standard industrial sensors, the 1600 architecture utilises additional internal strain gauges to heavily optimise accuracy performance and thermal characteristics.

It functions symmetrically across both tension and compression vectors within a single LowProfile unit, delivering an elite non-repeatability profile of just 0.005% alongside a minimal 0.01% creep rating. To completely eliminate thread engagement variables and poor repeatability at lower force limits, each cell arrives from the factory outfitted with a locked-in, factory-installed calibration adapter and an integrated high-precision base. Handcrafted with full eccentric load compensation to cancel out side-force or off-axis installation errors, the sensor operates across a wide capacity range from 500 lbf up to 200K lbf (2.2 kN to 900 kN). Units feature high output signals up to 4mV/V and integrated TEDS self-identification hardware, backed by a 3 run NIST-traceable ASTM E74.

  • Warranted performance to 0.02%
  • Warranted non-repeatability to +/- 0.005% or better
  • 9 capacities from 2.2kN to 900kN (500lbf to 200Klbf) High output to 4.0 mV/V for highest resolution.
  • Fitted with locked-in calibration adaptor
  • Moment compensation for enhanced insensitivity to extraneous load.
  • Excellent temperature characteristics
  • Low deflection
  • Individual 3-run multi-point calibration certificate traceable to National Standards
PARAMETERSMODEL
16101610161016201632, 1640
CAPACITY
Measuring RangeU.S. (lbf)5K, 10K1K, 2K50025K, 50K100K, 200K
Metric (kN)22, 454.5, 92.2110, 225450, 900
ACCURACY - (MAX ERROR)
Static Error Band - %FS±0.025±0.025±0.02±0.05±0.05
Nonlinearity - %FS±0.035±0.035±0.03±0.05±0.05
Hysteresis - %FS±0.045±0.035±0.02±0.05±0.05
Non-repeatability - %RO±0.005±0.005±0.005±0.005±0.005
Creep, in 20 min - %±0.01±0.01±0.01±0.01±0.01
Side Load Sensitivity - %±0.1±0.1±0.1±0.1±0.1
Eccentric Load Sensitivity - % / in±0.1±0.1±0.1±0.1±0.1
TEMPERATURE
Compensated Range°F+14 to +113+14 to +113+14 to +113+14 to +113+14 to +113
°C-10 to +45-10 to +45-10 to +45-10 to +45-10 to +45
Operating Range°F-67 to +194-67 to +194-67 to +194-67 to +194-67 to +194
°C-55 to +90-55 to +90-55 to +90-55 to +90-55 to +90
Effect on Zero - %RO MAX°F±0.0008±0.0008±0.0008±0.0008±0.0008
°C±0.0014±0.0014±0.0014±0.0014±0.0014
Effect on Output - %RO MAX°F±0.0008±0.0008±0.0008±0.0008±0.0008
°C±0.0014±0.0014±0.0014±0.0014±0.0014
ELECTRICAL
Rated Output - mV/V (Nominal)44244
Bridge Resistance - Ohm (Nominal)350350350350350
MECHANICAL
Safe Overload - %CAP±150±150±150±150±150
Deflection @ ROin0.0040.0040.0020.0060.01
mm0.1020.1020.0510.1520.254
Optional Base - P/N (Metric)installed_baseinstalled_baseinstalled_baseinstalled_baseinstalled_base
Weightlb23.583.858171
kg10.663.631.7226.3177.56
MaterialAluminumAluminumAluminumAlloy SteelAlloy Steel
CalibrationTension & Compression

  • Multiple independent bridge outputs
  • Special fittings
  • Integral cable or choice of connector(s) Mechanical adaptations
  • Standardised output
  • A2LA accredited calibration
  • ID Chip or TEDS (Transducer Electronic Data Sheet)
  • ISO/ASTM or MILSPEC calibration
  • Special thread sizes
  • Display electronics or data acquisition
  • Load cell with fitted base and locked-in calibration adaptor
  • Connector output versions supplied with mating part(s)
  • NIST traceable, ASTM E74 or ISO376 calibration certificate with plots and calculations